Use default impedance for load standard transmission line
This commit is contained in:
parent
069f68a5d0
commit
6aa1157992
|
|
@ -253,7 +253,7 @@ class Calkit::SOLT Calkit::toSOLT(double frequency, bool male_standards)
|
|||
auto imp_L = complex<double>(0, frequency * 2 * M_PI * Load.Lseries);
|
||||
imp_load += imp_L;
|
||||
ref.Load = (imp_load - complex<double>(50.0)) / (imp_load + complex<double>(50.0));
|
||||
ref.Load = addTransmissionLine(ref.Load, Load.Z0, Load.delay*1e-12, 0, frequency);
|
||||
ref.Load = addTransmissionLine(ref.Load, 50.0, Load.delay*1e-12, 0, frequency);
|
||||
}
|
||||
|
||||
if(Open.useMeasurements) {
|
||||
|
|
|
|||
Loading…